<?xml version='1.0' encoding='UTF-8'?><?xml-stylesheet href="http://www.blogger.com/styles/atom.css" type="text/css"?><feed xmlns='http://www.w3.org/2005/Atom' xmlns:openSearch='http://a9.com/-/spec/opensearchrss/1.0/' xmlns:blogger='http://schemas.google.com/blogger/2008' xmlns:georss='http://www.georss.org/georss' xmlns:gd="http://schemas.google.com/g/2005" xmlns:thr='http://purl.org/syndication/thread/1.0'><id>tag:blogger.com,1999:blog-4735270937248608841</id><updated>2024-11-01T03:36:20.087-07:00</updated><category term=".calibration of pressure gauge"/><title type='text'>instrumentation</title><subtitle type='html'>this blog is all about the calibration of instruments from temperature, flow, pressure and level</subtitle><link rel='http://schemas.google.com/g/2005#feed' type='application/atom+xml' href='http://instrumentation2010.blogspot.com/feeds/posts/default'/><link rel='self' type='application/atom+xml' href='http://www.blogger.com/feeds/4735270937248608841/posts/default?redirect=false'/><link rel='alternate' type='text/html' href='http://instrumentation2010.blogspot.com/'/><link rel='hub' href='http://pubsubhubbub.appspot.com/'/><author><name>ong-ong</name><uri>http://www.blogger.com/profile/09173291947273581406</uri><email>noreply@blogger.com</email><gd:image rel='http://schemas.google.com/g/2005#thumbnail' width='32' height='32' src='//blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEiqHPJNAg02XlusIJmHAzJXGf7mf38AjgH3C8SjP0y9URJWmddC_UtAKMTJEdZeZT9N_F54CXAX8nMru2mH4UjtUQrGlEn9zdb_nGq78wKO5Ea_nqnyWGbQVaVlhJYwqw/s220/Yosibaby000.gif'/></author><generator version='7.00' uri='http://www.blogger.com'>Blogger</generator><openSearch:totalResults>1</openSearch:totalResults><openSearch:startIndex>1</openSearch:startIndex><openSearch:itemsPerPage>25</openSearch:itemsPerPage><entry><id>tag:blogger.com,1999:blog-4735270937248608841.post-7556808621736618467</id><published>2010-07-12T06:56:00.002-07:00</published><updated>2010-07-12T09:42:31.703-07:00</updated><category scheme="http://www.blogger.com/atom/ns#" term=".calibration of pressure gauge"/><title type='text'>pressure gauge calibration using deadweight tester</title><content type='html'>&lt;a href=&quot;https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEhw_2kEq4GmOx8VZDJ5PpQdRBZNNHXnZc2jivStCQqatumSyMi1h49Zz_I3nDs8DJcVd4e-CuRg-0xaXHDQiAbH49WcLerfa1wGSmUOJZ2Iks1vkya2gtdyfq1sPd2Cw7MA6Me_9K7EHXY/s1600/DSC05855.JPG&quot; onblur=&quot;try {parent.deselectBloggerImageGracefully();} catch(e) {}&quot;&gt;&lt;img alt=&quot;&quot; border=&quot;0&quot; id=&quot;BLOGGER_PHOTO_ID_5493025749173203474&quot; src=&quot;https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEhw_2kEq4GmOx8VZDJ5PpQdRBZNNHXnZc2jivStCQqatumSyMi1h49Zz_I3nDs8DJcVd4e-CuRg-0xaXHDQiAbH49WcLerfa1wGSmUOJZ2Iks1vkya2gtdyfq1sPd2Cw7MA6Me_9K7EHXY/s320/DSC05855.JPG&quot; style=&quot;cursor: pointer; float: right; height: 240px; margin: 0pt 0pt 10px 10px; width: 320px;&quot; /&gt;&lt;/a&gt;&lt;br /&gt;
&lt;br /&gt;
STEPS:&lt;br /&gt;
&lt;br /&gt;
1. Set up the equipments needed for calibration&lt;br /&gt;
note: know first the maximum capacity of the dead weight tester whether it can acommodate the range of the gauge calibrated.&lt;br /&gt;
2. prepare the manual and date sheet that will be used for details of the instruments to be calibrated.&lt;br /&gt;
some information needed for calibration to be input to data sheet:&lt;br /&gt;
. equipment name/id no&lt;br /&gt;
.ambient temperature&lt;br /&gt;
. relative humidity&lt;br /&gt;
.date of calibration&lt;br /&gt;
.date of next calibration&lt;br /&gt;
3. wrap the gauge fitting of teflon tape in clockwise direction&lt;br /&gt;
4.place the gauge to be tested to the manifold assembly ( with gauge fitting match to the manifold joint) use correct size of wrench and make sure there is no leak.&lt;br /&gt;
5. if the gauge has to be tested has for example 0 to 100 psi range, use the 5 points scale as reading reference.&lt;br /&gt;
0% -low pressure ( zero reading)&lt;br /&gt;
.100% high pressure&lt;br /&gt;
(full scale reading)&lt;br /&gt;
here are the i.e of 5 points calibration using zero to 100&lt;br /&gt;
___________________________________________________&lt;br /&gt;
[percent_____!___equivalent psi____!________________________________&lt;br /&gt;
[___0%_____!____0____________!_____________________&lt;br /&gt;
[___25%____!_____25__________!______________________&lt;br /&gt;
[____50%___!____50___________!______________________________&lt;br /&gt;
[____75%___!____75___________!_______________________&lt;br /&gt;
[___100%___!____100__________!&lt;br /&gt;
--------------------------------------------------------------------------------&lt;br /&gt;
6. check the pointer intial reading 0 psi. If not, zero adjustment shall be made by adjusting the pinter zero.&lt;br /&gt;
7.Place weights on the cylinder corresponding to each point scale of the pressure gauge. increase the actuating pressure equivalent to 0%, 25%,50%,75% and 100% of the assigned range&lt;br /&gt;
8. Tightly close the release valve by turning clockwise.&lt;br /&gt;
9.Pump gently until it weights floats to an ideal level&lt;br /&gt;
10. rotate weights clockwise very gently and continuously while pumping.&lt;br /&gt;
11.If cylinder base exceeds above or below the ideal level,, rotate fine pressure adjustment lever.&lt;br /&gt;
12. record the pointer indication values for each point scale.&lt;br /&gt;
.upscale ( o%,25%,50%,75% and 100%)&lt;br /&gt;
.downscale ( 100%,75%,50%25% and 0%)&lt;br /&gt;
this is the &quot;as-found data&quot;.&lt;br /&gt;
13. if the &quot;as found&quot; data exceeds the required accuracy, calibration should be done. a zero or span adjustment must be made.&lt;br /&gt;
span = maximum range-minimum range&lt;br /&gt;
14. repeat the procedure until the readings for each point scale is within the specified limits of accuracy.&lt;br /&gt;
15.record the readings for each point scale and this is the as left date&lt;br /&gt;
16. when done, release the built system pressure by opening the release valve&lt;br /&gt;
17.put the calibrator sticker.&lt;br /&gt;
18.make the calibration certificate with authorize signatures.&lt;br /&gt;
19.housekeeping&lt;br /&gt;
&amp;nbsp;---------------&lt;br /&gt;
&lt;script src=&quot;http://www.assoc-amazon.com/s/link-enhancer?tag=widgetsamazon-20&amp;amp;o=1&quot; type=&quot;text/javascript&quot;&gt;&lt;/script&gt;</content><link rel='replies' type='application/atom+xml' href='http://instrumentation2010.blogspot.com/feeds/7556808621736618467/comments/default' title='Post Comments'/><link rel='replies' type='text/html' href='http://instrumentation2010.blogspot.com/2010/07/pressure-gauge-calibration-using.html#comment-form' title='4 Comments'/><link rel='edit' type='application/atom+xml' href='http://www.blogger.com/feeds/4735270937248608841/posts/default/7556808621736618467'/><link rel='self' type='application/atom+xml' href='http://www.blogger.com/feeds/4735270937248608841/posts/default/7556808621736618467'/><link rel='alternate' type='text/html' href='http://instrumentation2010.blogspot.com/2010/07/pressure-gauge-calibration-using.html' title='pressure gauge calibration using deadweight tester'/><author><name>ong-ong</name><uri>http://www.blogger.com/profile/09173291947273581406</uri><email>noreply@blogger.com</email><gd:image rel='http://schemas.google.com/g/2005#thumbnail' width='32' height='32' src='//blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEiqHPJNAg02XlusIJmHAzJXGf7mf38AjgH3C8SjP0y9URJWmddC_UtAKMTJEdZeZT9N_F54CXAX8nMru2mH4UjtUQrGlEn9zdb_nGq78wKO5Ea_nqnyWGbQVaVlhJYwqw/s220/Yosibaby000.gif'/></author><media:thumbnail xmlns:media="http://search.yahoo.com/mrss/" url="https://blogger.googleusercontent.com/img/b/R29vZ2xl/AVvXsEhw_2kEq4GmOx8VZDJ5PpQdRBZNNHXnZc2jivStCQqatumSyMi1h49Zz_I3nDs8DJcVd4e-CuRg-0xaXHDQiAbH49WcLerfa1wGSmUOJZ2Iks1vkya2gtdyfq1sPd2Cw7MA6Me_9K7EHXY/s72-c/DSC05855.JPG" height="72" width="72"/><thr:total>4</thr:total></entry></feed>